Choose Lonnmeter for accurate and intelligent measurement!

LONN-700CM Inline Tuning Fork Density & Concentration Meter

The LONN-700CM inline tuning fork density meter is a high-precision online density meter designed for continuous, real-time liquid density measurement and concentration monitoring in industrial processes. Engineered with a robust vibrating fork sensor, this tuning fork density meter delivers stable, accurate density readings across demanding applications — from corrosive chemicals in petrochemical plants to sugar solutions in food production lines.

Unlike bulky coriolis mass flow systems, the LONN-700CM vibrating fork density sensor installs directly into your pipeline via flange connection, requiring minimal straight pipe run and offering straightforward integration with existing DCS and PLC systems via 4-20mA output or RS485 Modbus protocol.

Key Benefits:

  • ✅ High Accuracy: ±0.002 g/cm³ density precision
  • ✅ Wide Range: 0-2 g/cm³ (0-2000 kg/m³) continuous inline density measurement
  • ✅ Temperature Resilient: -25°C to +120°C standard, up to +150°C for high temperature process applications
  • ✅ Pressure Rated: Up to 2.5 MPa standard, 6.3 MPa for demanding pipeline conditions
  • ✅ Multi-Output: 4-20mA + RS485 Modbus — standard process industry signal formats
  • ✅ Hazardous Area Ready: Explosion proof inline density meter configuration (ATEX available)

How It Works — Tuning Fork Density Meter Principle

The LONN-700CM inline tuning fork density meter operates on a proven vibrating fork principle: an electronic circuit drives a pair of metal tuning forks to vibrate at their natural resonant frequency submerged in the process liquid. As the liquid density changes, the vibrational frequency shifts proportionally — denser liquids damp the vibration, while lower-density fluids allow faster oscillation. The instrument’s built-in temperature sensor applies automatic 20 ppm/°C compensation, calculating the corrected density value referenced to 25°C.

This continuous density measurement approach offers significant advantages over alternative technologies. Compared to an ultrasonic density analyzer, the tuning fork sensor performs better in clean, low-particulate liquids. Compared to a nuclear or radiometric densitometer, the LONN-700CM non nuclear density meter eliminates radioactive source licensing, safety inspections, and complex compliance procedures — making it the preferred solution for mines, chemical plants, and food facilities seeking a safer, lower-maintenance continuous density monitoring system.

A coriolis mass flow system achieves higher theoretical accuracy but demands more complex installation, larger pipe runs, and significantly higher cost. The LONN-700CM vibrating fork density sensor provides an ideal balance: solid industrial accuracy, easy inline installation, and a competitive price point backed by LONNMETER’s direct-from-factory guarantee.


Technical Specifications

Parameter Specification
Measurement Principle Tuning fork vibration frequency
Density Range 0 ~ 2 g/cm³ (0 ~ 2000 kg/m³)
Density Accuracy ±0.002 g/cm³
Density Repeatability ±0.0001 g/cm³
Concentration Range 0 ~ 100%
Concentration Accuracy ±0.25%
Temperature Coefficient 20 ppm/°C (after calibration)
Process Temperature Standard: -25°C ~ +120°C; Custom: -25°C ~ +150°C
Max Working Pressure Standard: 2.5 MPa; Custom: 6.3 MPa
Wetted Materials 316L SS, Hastelloy, Titanium, Zirconium (optional)
Fork Surface Treatment Standard polish / PTFE coating / Electro-polish
Enclosure Rating IP65, Aluminum alloy
Output Signals 4-20mA + RS485 (Modbus RTU)
Max Medium Viscosity 2000 cP
Particle Size Limit <10 μm (recommended); 10-50 μm (limited use)
Max Suspended Solids ≤40%
Max Gas Content ≤20%
Power Supply 24V DC
Display LCD with backlight
Mounting Flange connection (ANSI/DIN/JIS)

Applications by Industry

Petrochemical & Oil Refining

This inline density meter handles crude oil density monitoring in pipeline transfer stations and fuel oil density tracking at refinery injection points. The Hastelloy-wetted version serves sulfuric acid concentration monitoring in alkylation units. For fuel quality control across the processing chain, continuous inline density measurement ensures consistent product quality and batch control.

Chemical Processing

The LONN-700CM is specified as a process density meter for hydrochloric acid, sodium hydroxide (caustic soda), methanol, ethanol, glycerin, and hydrogen peroxide processes. The Hastelloy or Zirconium material option provides reliable corrosive liquid density monitoring where aggressive media would damage standard stainless steel sensors.

Pharmaceutical & Food & Beverage

For sugar concentration measurement, the LONN-700CM delivers precise brix monitoring in sugar refineries and beverage production. Sanitary-grade wetted parts make this a sanitary process density sensor for syrups, fruit juice concentration monitoring, and brewing process control. As a pharmaceutical process instrument, it monitors solvent density in alcohol extraction and API concentration processes.

Battery Manufacturing

The inline density meter configuration tracks sulfuric acid concentration in lead-acid battery formation and monitors lithium hydroxide concentration in electrolyte preparation — a critical quality control parameter for battery manufacturers.

Mining & Mineral Processing

Though optimized for cleaner media, the LONN-700CM handles light slurry concentration monitoring in pre-concentration circuits with careful particle sizing. For heavy slurry applications, consider the LONN-7000 series ultrasonic concentration meter as an alternative.

Environmental & Waste Treatment

The online density meter monitors wastewater concentration in MVR evaporator concentrate streams and provides slurry density analysis for limestone-gypsum slurry in power plant scrubbers.


Installation & Integration

Physical Installation

  • Mounting Position: Vertical or horizontal pipeline (avoid gas pockets)
  • Flange Standards: ANSI 150#-600#, DIN PN16-PN64, JIS 10K-40K
  • Straight Pipe Requirements: 5D upstream, 3D downstream (D = pipe diameter)
  • Process Connection: Flange mounting with gasket

Electrical Connections

  • Power: 24V DC (two-wire loop-powered)
  • Analog Output: 4-20mA (density or concentration selectable)
  • Digital Output: RS485 Modbus RTU
  • Cable Entry: M20×1.5 or ½” NPT

DCS/PLC Integration

  • Standard 4-20mA input card compatible
  • Modbus RTU register map provided
  • HART protocol (optional)

How to Install

  1. Select a pipeline location with sufficient straight run (5D upstream minimum)
  2. Ensure the process flanges match your pipe specification (ANSI/DIN/JIS)
  3. Install with the tuning fork sensors oriented perpendicular to flow
  4. Connect power (24V DC) and signal wiring (4-20mA and/or RS485)
  5. Configure the 4-20mA output scaling via the local LCD display
  6. Perform initial verification with a known reference fluid

Density Meter Comparison

Feature Tuning Fork (LONN-700CM) Ultrasonic Density Analyzer Nuclear / Radiometric Coriolis Mass Flow
Density Accuracy ±0.002 g/cm³ ±0.005 g/cm³ ±0.001 g/cm³ ±0.0001 g/cm³
Installation Inline flange Inline / Clamp-on Radiation shield Large supports
Maintenance Virtually maintenance-free Low High (radiation compliance) Periodic tube cleaning
Safety No radiation hazard No radiation hazard Radioactive source regulated No radiation hazard
Cost Medium Medium High Very High
Best For Clean liquids, chemicals, food Slurries, opaque liquids Mining, dredging, heavy slurry Ultimate precision, custody transfer

Density Meter Selection Guide

Choose the LONN-700CM if:

  • You need a process density meter for clean or low-particulate industrial liquids
  • Your application requires inline density measurement with 4-20mA or RS485 output
  • You want a non nuclear density meter to avoid radioactive source compliance
  • You need a high temperature density meter or high pressure process configuration
  • Your budget needs an Emerson density meter alternative or Endress+Hauser density meter alternative at competitive factory-direct pricing

Consider alternative technologies if:

  • Your process contains high-concentration slurries with abrasive solids → LONN-7000 series (ultrasonic)
  • You require laboratory-level custody transfer precision → Coriolis mass flow meter
  • Your application involves radioactive environment monitoring → Berthold nuclear system (with full compliance infrastructure)

FAQs

Q: How does a tuning fork density meter work?
A: The LONN-700CM operates on a proven vibrating fork principle: an electronic circuit drives a pair of metal tuning forks to vibrate at their natural resonant frequency submerged in the process liquid. As the liquid density changes, the vibrational frequency shifts with the density — this relationship is calibrated and temperature-compensated to produce accurate density readings in real time.

Q: What is the difference between a tuning fork density sensor and an ultrasonic density analyzer?
A: A tuning fork density sensor measures the damping effect of liquid on a vibrating fork, ideal for clean liquids. An ultrasonic density analyzer uses sound wave propagation speed, better suited for slurries and opaque media with higher suspended solids.

Q: Is the LONN-700CM a non nuclear density meter? Can it replace a nuclear densitometer?
A: Yes. The LONN-700CM is a non nuclear density meter that uses mechanical vibration technology — no radioactive source is required. It is a direct alternative for facilities looking to replace Berthold or other nuclear/radiometric density meters while eliminating radiation licensing and safety compliance burden.

Q: How do I calibrate the inline density meter?
A: The unit ships factory-calibrated. For field verification, immerse the tuning fork in a reference liquid of known density (e.g., distilled water = 1.000 g/cm³ at 25°C) and adjust the calibration offset via the LCD menu. We recommend annual recalibration for critical process applications.

Q: Can this online density meter measure concentration directly?
A: Yes. When the density-to-concentration relationship is known (e.g., sulfuric acid solutions, sugar/water Brix), we can pre-program the concentration curve so the outputs display directly as concentration %. Provide your media type and target concentration range when requesting a quote.


Request for Quote

Ready to get pricing? Fill out our RFQ form with the following information:

  1. Application (liquid type, industry)
  2. Target density range or concentration range
  3. Process temperature and pressure
  4. Pipe size and flange standard
  5. Required outputs (4-20mA, Modbus, HART)
  6. Quantity required

Response time: Within 24 hours for standard configurations.

As a trusted process density meter manufacturer, LONNMETER supplies factory-direct pricing with full technical documentation — including detailed product datasheet, full specification sheet, and Modbus register map for every unit. The LONN-700CM Inline Tuning Fork Density & Concentration Meter exemplifies advanced technology in liquid density measurement. This industrial density meter operates as a vibrating fork density meter, providing precise and continuous density readings essential for various applications. With its robust density sensor, this non-nuclear density meter ensures reliable performance in challenging environments. Designed for seamless integration, the inline density meter serves as an effective density analyzer, enabling real-time monitoring and enhanced process efficiency.

Leave Your Message

Request for Quotation

Please provide your application parameters for a technical assessment and formal quote.